Comparative Effects of Thermocycling and Artificial Saliva Immersion on Surface Roughness and Light Transmittance of 3D-Printed and Thermoformed Orthodontic Aligners: An In-Vitro Study

Document Type : Original Article

Authors
1 Post graduate Researcher, Department of Orthodontics, College of Dentistry, University of Baghdad, Baghdad, Iraq
2 Assistant Professor, Department of Orthodontics, College of Dentistry, University of Baghdad, Baghdad, Iraq
Abstract
Aim
This in vitro study compared the effects of thermocycling and artificial saliva immersion on the light transmittance and surface roughness of thermoformed (TF) and directly 3D-printed (DP) orthodontic aligners.
Methods
Twenty-four aligners were used: 12 multilayer thermoplastic aligners (CA® Pro+, Scheu Dental) and 12 directly 3D-printed resin aligners (Tera Harz TC-85, Graphy). Each material was divided into three independent subgroups of four specimens: as-received (AR), after 500 thermocycles (5 °C to 55 °C), and after 14 days of immersion in artificial saliva at 37 °C (AS). Light transmittance (T%) was measured with a UV/Vis spectrophotometer over 400–700 nm, and surface roughness (arithmetic mean height, Sa) was assessed by atomic force microscopy over a 2 × 2 µm area. Between-material comparisons used independent-samples t-test, and comparisons among the three conditions within each material used one-way ANOVA with Tukey post hoc tests (α = 0.05).
Results
Baseline T% did not differ significantly between TF and DP aligners (95.79 ± 1.94% vs. 87.26 ± 7.38%;
P = 0.067). After thermocycling, T% was 52.37 ± 19.23% for TF and 10.49 ± 4.48% for DP aligners
(P = 0.005). After saliva immersion, T% was 65.33 ± 5.83% for TF and 12.45 ± 7.35% for DP aligners
(P < 0.001). Sa after thermocycling was 23.16 ± 13.15 nm for TF and 70.99 ± 18.12 nm for DP aligners
(P = 0.005). After saliva immersion, Sa was 25.44 ± 23.15 nm for TF and 52.47 ± 5.51 nm for DP aligners
(P = 0.063). Within TF aligners, T% (P = 0.001) and Sa (P = 0.02) differed among conditions. Within DP aligners, T% differed (P < 0.001), but Sa did not (P = 0.14). All mean Sa values were below 500 nm.

Conclusion
Under these short-term aging conditions, thermoformed (TF) aligners had higher light transmittance and lower surface roughness than directly 3D-printed (DP) aligners after thermocycling. Because each aging condition was applied to an independent subgroup of four specimens, the sample is small, and aligner thickness differed between materials, these findings are exploratory and should be interpreted with caution.
Keywords
Subjects

1. Abbing A, Koretsi V, Eliades T, Papageorgiou SN. Duration of orthodontic treatment with fixed appliances in adolescents and adults: a systematic review with meta-analysis. Prog Orthod. 2020;21(1):37. doi: 10.1186/s40510-020-00334-4.
2. Can E, Panayi N, Polychronis G, Papageorgiou SN, Zinelis S, Eliades G, Eliades T. In-house 3D-printed aligners: effect of in vivo ageing on mechanical properties. Eur J Orthod. 2022;44(1):51-55. doi: 10.1093/ejo/cjab022.
3. Neoh SP, Khantachawana A, Chintavalakorn R, Santiwong P, Srikhirin T. Comparison of physical, mechanical, and optical properties between thermoplastic materials and 3-dimensional printing resins for orthodontic clear retainers. Am J OrthodDentofacialOrthop.2025;167(1):95-109.e1.doi: 10.1016/j.ajodo.2024.06.010.
4. Wu C, Mangal U, Bai N, Kim H, Hwang G, Cha JY, Lee KJ, Kwon JS, Choi SH. Effects of printing layer thickness and build orientation on the mechanical properties and color stability of 3D-printed clear aligners. Angle Orthod. 2025;95(4):355-361. doi: 10.2319/081624-672.1.
5. Cremonini F, Vianello M, Bianchi A, Lombardo L. A spectrophotometry evaluation of clear aligners transparency: comparison of 3D-printers and thermoforming disks in different combinations. applied sciences.2022;12(23):11964. https://doi.org/10.3390/app122311964
6. Porojan L, Vasiliu RD, Porojan SD, Bîrdeanu MI. Surface quality evaluation of Removable Thermoplastic Dental Appliances Related to Staining Beverages and cleaning agents. Polymers (Basel). 2020;12(8):1736.doi: 10.3390/polym12081736.
7. Eslami S, Kopp S, Goteni M, Dahmer I, Sayahpour B. Alterations in the surface roughness and porosity parameters of directly printed and Invisalign aligners after 1 week of intraoral usage: An in vivo prospective investigation. Am J Orthod Dentofacial Orthop. 2024;165(1):73-79. doi: 10.1016/j.ajodo.2023.07.013.
8. Šimunović L, Pečanić P, Marić AJ, Haramina T, Rakić IŠ, Meštrović S. Impact of various cleaning protocols on the physical and aesthetic properties of 3D-printed orthodontic aligners. Sci Rep. 2025;15(1):19022. doi: 10.1038/s41598-025-04096-z.
9. Albilali AT, Baras BH, Aldosari MA. Evaluation of mechanical properties of different thermoplastic orthodontic retainer materials after thermoforming and thermocycling. Polymers (Basel). 2023;15(7):1610. doi: 10.3390/polym15071610.
10. Kuntz L, Aranda L, Rapin C, Canceill T, Vande Vannet B, Fawaz P. Effects of aging on the tensile strength and surface condition of orthodontic aligners: a comparative study of five models. Eur J Orthod. 2024;46(6):cjae063. doi: 10.1093/ejo/cjae063.
11. Koletsi D, Panayi N, Laspos C, Athanasiou AE, Zinelis S, Eliades T. In vivo aging-induced surface roughness alterations of Invisalign® and 3D-printed aligners. J Orthod. 2023;50(4):352-360. doi: 10.1177/14653125221145948.
12. Šimunović L, Čekalović Agović S, Marić AJ, Bačić I, Klarić E, Uribe F, Meštrović S. Color and chemical stability of 3D-printed and thermoformed polyurethane-based aligners. Polymers (Basel). 2024;16(8):1067. doi: 10.3390/polym16081067.
13. Tartaglia GM, Mapelli A, Maspero C, Santaniello T, Serafin M, Farronato M, Caprioglio A. Direct 3D printing of clear orthodontic aligners: current state and future possibilities. Materials (Basel). 2021;14(7):1799. doi: 10.3390/ma14071799.
14. Fidan M, Gelgor IE. Effect of aging and mechanical brushing on color stability and translucency of three-dimensionally printed and thermoformed aligners of different thicknesses. Angle Orthod. 2026;96(1):83-92. doi:10.2319/011325-50.1.
 

Articles in Press, Accepted Manuscript
Available Online from 29 November 2026